Features |
| Pressure type: | Gauge pressure | |
| Measuring range: | 0 bar ... 10 bar | |
| Process temperature: | -20 °C ... 150 °C | |
| CE-conformity: | Yes | |
| Output signal: | 4...20 mA | |
Performance |
| Accuracy: | <= +/- 0.25 % (of span) | |
| Non-linearity: | <= +/- 0.2 % (of span (Best Fit Straight Line, BFSL) according to IEC 61298-2) | |
| Long-term drift/one-year stability: | ≤- 0.2 % of span | |
| Temperature coefficient in rated temperature range: | Mean TC of span: ≤- 0.2 % of span / 10 K, Mean TC of zero: ≤- 0.2 % of span/ 10 K, with pressure ranges 0 bar ... 0.6 bar to 0 bar ... 25 bar, Mean TC of zero: ≤- 0.25 % of span/ 10 K with pressure range 0 bar ... 0.4 bar≤- 0.25 % of span/ 10 K, with pressure range 0 bar ... 0.4 bar, Mean TC of zero: ≤- 0.4 % of span/ 10 K with pressure range 0 bar ... 0.25 bar | |
| Rated temperature range: | 0 °C ... 80 °C | |
| Non-repeatability: | <= +/- 0.1 % of span | |
| Response time (10 % ... 90 %): | ≤- 10 ms | |
Mechanics/electronics |
| Process connection: | Conical coupling (DIN 11851) DN 50 with union nut | |
| Wetted parts: | Stainless steel 1.4435 | |
| Housing material: | Stainless steel 1.4571 | |
| Enclosure rating: | IP 67 | |
| Connection type: | Field housing IP 67 | |
| Supply voltage: | 11 V DC ... 30 V DC | |
| Electrical safety: | Short-circuit protection: QA towards M, Protection class: III, Overvoltage protection: 36 V DC, Reverse polarity protection: L+ towards M | |
| Dielectric strength: | 500 V DC, NEC Class 02 power supply (low voltage and low current max. 100 VA even under fault conditions) | |
| Protection class III: | Yes | |
| Weight sensor: | 500 g | |
| CE-conformity: | EMC directive: 2004/108/EC, EN 61 326-2-3 | |
Ambient data |
| Storage temperature: | -40 °C ... 100 °C | |
| Ambient temperature: | -20 °C ... 80 °C | |
| Shock load: | 500 g according to IEC 60068-2-27 (mechanical shock) | |
| Vibration load: | 15g per IEC 60068-2-6 (vibration under resonance) | |
*изображение на сайте может отличаться от реального