Analyzer systems | |||||||||||||||||||||||||||||||||||||||||||||
| |||||||||||||||||||||||||||||||||||||||||||||
Technical data | |||
Measuring principle: | Gas filter correlation, Interference filter correlation, Zirconium dioxide sensor | ||
Number of measuring components: | 6 | ||
Measuring ranges: | CO: 0 ... 800 ppm / 0 ... 5 Vol.-% CO2: 0 ... 20 Vol.-% / 0 ... 100 Vol.-% H2O: 0 ... 1 Vol.-% / 0 ... 60 Vol.-% HCl: 0 ... 700 ppm / 0 ... 12,000 ppm N2O: 0 ... 500 ppm / 0 ... 7,500 ppm NH3: 0 ... 530 ppm / 0 ... 5,300 ppm NO: 0 ... 750 ppm / 0 ... 15,000 ppm NO2: 0 ... 500 ppm / 0 ... 4,000 ppm O2: 0 ... 1 Vol.-% / 0 ... 21 Vol.-% SO2: 0 ... 200 ppm / 0 ... 10 Vol.-% | ||
Remark: | 2 measuring ranges per component, Other measuring ranges and components on request | ||
Sample quantity: | 150 ... 800 l/h | ||
Process temperature: | Sample gas inlet: ≤-220 °C, Process: ≤-1,300 °C | ||
Process pressure: | 800 ... 1,200 hPa | ||
Ambient temperature: | 5 ... 35 °C temperature change max. ±10 °C/h, With cooling device: 5 ... 45 °C temperature change max. ±10 °C/h | ||
Ambient humidity: | ≤-80 % non-condensing | ||
Electrical safety: | CE | ||
Enclosure rating: | IP 43 higher enclosure rating on request | ||
Analogue outputs: | 2 outputs: 0/4 ... 22 mA, 500 Ω- electrically isolated- additional outputs if using I/O modules (option) | ||
Analogue inputs: | 2 inputs: 0/4 ... 22 mA, 100 Ω- electrically isolated- additional inputs if using I/O modules (option) | ||
Digital outputs: | 2 power relays electrically isolated- additional outputs if using I/O modules (option), 3 floating contacts additional outputs if using I/O modules | ||
Digital inputs: | 2 floating contacts additional inputs if using I/O modules | ||
Interfaces: | Ethernet | ||
Bus protocol: | Modbus TCP, OPC | ||
Operation: | Application specific programs for system control, Via operating unit in the analyzer, Via software SOPAS ET, Two operating levels, one password-protected | ||
Device types: | Steel sheet cabinet | ||
System components: | Analyzer cabinet, Heated sample gas line, Sampling probe, Sample point switching (max. 8 sample points) | ||
Correction functions: | Internal adjustment unit (option) | ||
Test functions: | Automatic control cycle for zero and span point, Status LEDs "Power", "Maintenance" and "Fault" | ||
Note: | The scope of delivery depends on application and customer specifications. |